To use alongside the PRESTO temperature controllers, the Julabo VFCpro units measure flow rates and calculate calorimetric values. Determine with precision the kinetic and dynamics of reactions and help scale-up of processes from laboratory quantities to pilot and production volumes.
Measurements are shown directly on the PRESTO display.
Together the VFCpro and PRESTO control measure flow rate with an accuracy of 0.8 to 1.5%. This corresponds to a flow of approximately 0.4 to 0.75 L/m.
Calorimetric measurement accuracy depends on the correct specification of heat capacities of bath fluids. The specifications for the bath fluids Thermal HL40, Thermal HL45 and Thermal HL80 are stored in the PRESTO settings, along with user inputted specifications.
Calorimetric measurement accuracy also depends on the quality of sensor taring (0 set) at the beginning of each measurement, the absolute introduced energy, and the mass flow of bath fluid. The VFCpro-24 requires a flow of at least 10L/m while the VFCpro-38 required 20L/m.
Features of the VFCpro Flow Measurement and Control:
Determine flow rate through an apparatus
Real-time measurement of heating and cooling capacity (kW)
Real-time measurement of flow rate (L/min)
Real-time measurement of calorimetric work (kWh)
Flow rate control for maintaining constant flow while controlling temperature
Working temperature range -100 to +300°C
Specifications (Click here):
Flow Measurement, Julabo VFCpro and VFC
Specifications
Model:
VFCpro-24
VFCpro-38
VFC Flow Measurement
Working temperature range: (°C)
-100 ≤ +300
-100 ≤ +300
-100 ≤ +300
Display:
Uses PRESTO display
Uses PRESTO display
-
Flow rate measurement accuracy: (%)
0.8 to 1.5
0.8 to 1.5
-
Flow rate measurement accuracy: (L/m)
0.4 to 0.75
0.4 to 0.75
-
Flow rate maximum: ( L/m)
50
100
50
Accuracy of calorimetric measurements:
Dynamic error contribution depending on bath fluid settings, sensor taring, absolute introduced energy and bath fluid mass flow
Dynamic error contribution depending on bath fluid settings, sensor taring, absolute introduced energy and bath fluid mass flow
Julabo VFCpro-24 Flow Measurement and Control, -100 to +300°C, maximum flow rate 50L/m, suitable for PRESTO temperature controllers A40, A80, A80t, W40, W80, W80t
Inquire
8980764
Julabo VFCpro-38 Flow Measurement and Control, -100 to +300°C, maximum flow rate 100L/m, suitable for PRESTO temperature controllers W91, W91t, W91tt, W91ttx, W91tx. W91x. W92, W92, W92t, W92tt, W92ttx. W92tx, W92x.
Inquire
8980782
Julabo VFC Flow Measurement, -100 to +30°C, maximum flow rate 50L/m