The Memmert Constant Climate Chambers Models HPP110, HPP260 and HPP750 are designed for long term stability testing for pharmaceutical products in accordance with ICH (International Conference on Harmonization of Technical Requirements for Registration of Pharmaceuticals for Human Use) standards and for any other application that requires testing under conditions of long term, precisely controlled environmental simulation.
The Peltier technology used in the HPP Constant Climate Chambers provides heating, cooling, humidifying and dehumidifying quietly, with little vibration, in a compact appliance that conserves energy and is a much more acceptable long term laboratory companion than a compressor based appliance.
The temperature control from +5 ºC to +70 ºC, as well as the active humidification and dehumidification from 10% to 90% relative humidity, are optimally tuned to the requirements of stability tests for the manufacturers of pharmaceutical products.
The quietness of the HPP Constant Climate Chambers make them ideal for breeding insects and reptiles.
With the choice of an optional light module the HPP Constant Climate Chambers are well suited for applications such as seed germination, plant cultivation, or insect breeding, especially Drosophila breeding, under constant climate and daylight conditions. A choice of cold white light (5500 Kelvin) or a combination of cold white and warm light (2700 Kelvin), dimmable and with average intensity of 10000 Lux, is available, using energy efficient LED emitters.
Double doors are standard, an exterior door of stainless steel and an interior one of glass. This feature reduces the potential for contamination and drops or rises in temperature, yet gives an optimal view of a sensitive load through wide interior glass doors
A wide range of options for programming and documentation using interfaces are available as part of the Memmert Generation 2012 class of products. See IPP Peltier-cooled Incubators, ICP compressor cooled incubators, or Universal Ovens for ordering information on Generation 2012 appliance options.
As much function as necessary, as much individuality as possible! Contact Rose Scientific to custom design the Memmert HPP Constant Climate Chamber that does what it needs to do with the resources available.
The ordering number is assembled from the catalog numbers of the various options at the time the HPP Constant Climate Chamber is ordered.
Memmert supplies a three-year guarantee for appliances, except for wear and tear parts
Features of the HPP Constant Climate Chambers:
Temperature range up to +70 ℃
Digitally controlled active humidification and dehumidification up to 90%
Three compact model sizes (108L, 256L, 749L)
Standard double doors reduce the possibility of contamination and drops or rises in temperature, and at the same time allow an optimal view through wide-area interior glass doors
Generation 2012 TwinDisplay technology
Peltier heating for efficient, stable, precise and very economic long term tests
Extremely quiet operation
Choose internal light options for germination, plant cultivation or insect breeding applications
Fault analysis for temperature and humidity control:
Self-diagnostics
Self-diagnostics
Self-diagnostics
Parameters adjustable on ControlCockpit:
Temperature, relative humidity, program time, time zones, season
Temperature, relative humidity, program time, time zones, season
Temperature, relative humidity, program time, time zones, season
Calibration facility, number of temperature values:
3
3
3
Calibration facility, number of relative humidity values: (%)
Two, fixed at 20 and 90
Two, fixed at 20 and 90
Two, fixed at 20 and 90
AtmoControl software on USB stick for creating, managing and transferring programs:
Yes
Yes
Yes
Temperature range, without optional light: (°C)
0 to +70
0 to +70
0 to +70
Temperature range with optional light: (°C)
+15 to +40
+15 to +40
-
Temperature display resolution for setpoint and actual values: (°C)
0.1
0.1
0.1
Relative humidity display resolution: (%)
0.1
0.1
0.1
Relative humidity setting accuracy: (%)
0.5
0.5
0.5
Individual control of upper and lower heating or cooling Peltier elements:
Yes
Yes
Yes
HeatBalance function to adjust upper and lower heating circuits: (%)
-
Yes, ± 50
Yes, ± 50
Over temperature protection:
Triple protection. Including one mechanical
Triple protection. Including one mechanical
Triple protection. Including one mechanical
Under temperature protection:
Digitally adjustable, Memmert ASF sounds an alarm
Digitally adjustable, Memmert ASF sounds an alarm
Digitally adjustable, Memmert ASF sounds an alarm
Number of Pt100 temperature sensors:
2
2
2
Humidity supply:
Distilled water from external tank by self-priming pump
Distilled water from external tank by self-priming pump
Distilled water from external tank by self-priming pump
Humidification:
Hot steam generator
Hot steam generator
Hot steam generator
Dehumidification:
Cold trap using the Peltier technology
Cold trap using the Peltier technology
Cold trap using the Peltier technology
Active humidifying and dehumidifying range: (%rh)
10 to 90
10 to 90
10 to 90
Programmable digital timer: (days:m)
0:01 to 99:00
0:01 to 99:00
0:01 to 99:00
Process time start:
When set temperature is reached
When set temperature is reached
When set temperature is reached
Maximum loading per grid: (kg)
30
30
30
Maximum loading of chamber: (kg)
150
200
200
Maximum number of shelves:
5
9
14
Stainless steel grid supplied:
2
2
2
Internal dimensions: W×D×H (mm)
560×400×480
640×500×800
1040×600×1200
Exterior dimensions: W×D×H (mm)
745×674×864
824×774×1183
1224×874×1726
Supplied on lockable castors:
No
No
Yes
Program stored in case of power failure:
Yes
Yes
Yes
Power: (V, Hz,)
115, 50/60 230, 50/60
115, 50/60 230, 50/60
115, 50/60 230, 50/60
Weight: (kg)
80
122
208
Number
Description
Price $
HPP110
Peltier Cooled Constant Climate Chamber HPP110 Generation 2012, Two High Definition TFT Color Displays, 108L, 115V, exterior dimensions (WxDxH) 745×674×864mm, interior dimensions 560×400×480mm, 80kg, 2 shelves provided, 5 shelves maximum
Inquire
HPP260
Peltier Cooled Constant Climate Chamber IPP260 Generation 2012, Two High Definition TFT Color Displays, 256L, 115V, exterior dimensions (WxDxH) 824×774×1183mm, interior dimensions 640×500×800mm, 122kg, 2 shelves provided, 9 shelves maximum
Inquire
HPP750
Peltier Cooled Constant Climate Chamber IPP750 Generation 2012, Two High Definition TFT Color Displays, 749L, 115V, exterior dimensions (WxDxH) 1224×874×1726mm, interior dimensions 1040×600×1200, 208kg, 2 shelves provided, 14 shelves maximum
Inquire
K1
Chamber modification for the application of reinforced perforated stainless steel shelves or stainless steel grids (bearing rails mounted in the working chamber), includes replacement of 2 standard grids by 2 reinforced grids, model sizes 450 and 750 only
Inquire
T7
Cold white 5500 Kelvin light module, light strips arranged on the side walls of the interior, 10 strips for model 110, 14 for models 260 and 750, illumination strength 10000 Lux, program controlled dimming from 0 to 100 % (in 10 % steps), ramp programming in combination with temperature and humidity
Inquire
T8
Cold white 5500 Kelvin plus warm white 2700 Kelvin light module, 10 LED light strips (5 alternating cold white light strips and 5 warm white light strips) arranged on the side wall of the interior, illumination strength 10000 LUX, program controlled dimming from 0 to 100% in 10 % steps, ramp programming in combination with temperature and humidity
Inquire
R3
Interior socket option, switches off using On/Off switch but not individually, moisture tight
Inquire
FX
Entry port option, 23mm Ø, for introducing connections at the side, moisture tight, can be closed by flap and silicone stopper, position F0 for left center center, F1 for left center top, F2 for right center center, F3 for right center top, F4 for special position left side, F5 for special position right side
Inquire
F6
Entry port option, 23mm Ø, for introducing connections at the back, moisture tight, can be closed by flap and silicon stopper, special position back side
Inquire
F7-Si
Entry port option, 40mm Ø, for introducing connections at the back, moisture tight, can be closed by silicon stopper, F7-Si special position back side, for HPP260 and HPP750
Inquire
V3-HPP
0-20 mA current loop interface option (-10 to +80°C corresponds to 4-20 mA) temperature controller actual value
Inquire
V6-HPP
0-20 mA current loop interface option (-10 to +80°C corresponds to 4-20 mA) temperature of a Pt100 sensor positioned flexibly in chamber (maximum 3 in TwinDisplay)
Inquire
V4-HPP
0-20 mA current loop interface option (-10 to +80°C corresponds to 4-20 mA) humidity controller actual value (0 to 100% rh corresponds to 4-20 mA)
Inquire
D00105
Works calibration certificate option for one pair of temperature and humidity values
Inquire
E20165
Stainless Steel Grid shelving (standard equipment) for HPP110
Inquire
E28891
Stainless Steel Grid shelving (standard equipment) for HPP260
Inquire
E20182
Stainless Steel Grid shelving (standard equipment) for HPP750
Inquire
E29767
Reinforced stainless steel grid shelving, maximum loading 60 kg, model sizes 110 and 160
Inquire
E29766
Reinforced stainless steel grid shelving, maximum loading 60 kg, model size 260
Inquire
B32190
Reinforced stainless steel grid shelving, maximum loading 60 kg, with guide bars and fixing screws, used only in connection with option K1, model sizes 450 and 750
Inquire
B00325
Perforated stainless steel shelf, 10 mm Ø holes, for HPP110
Inquire
B29725
Perforated stainless steel shelf, 10 mm Ø holes, for HPP260
Inquire
B34727
Perforated stainless steel shelf, 10 mm Ø holes, for HPP750
Inquire
B32191
Reinforced perforated stainless steel shelves, maximum loading 60 kg, with guide bars and fixing screws, used only in connection with option K1, for HPP750
Inquire
E02073
Slide in stainless steel drip tray, 15mm rim, may affect temperature distribution, cannot be used with option K1, for model sizes 110 and 160
Inquire
E29726
Slide in stainless steel drip tray, 15mm rim, may affect temperature distribution, cannot be used with option K1, for model size 260
Inquire
B04359
Bottom stainless steel drip tray, 15mm rim, may affect temperature distribution, cannot be used with option K1, for model sizes 110 and 160
Inquire
B29722
Bottom stainless steel drip tray, 15mm rim, may affect temperature distribution, cannot be used with option K1, for model size 260
Inquire
B04362
Bottom stainless steel drip tray, 15mm rim, may affect temperature distribution, cannot be used with option K1, for model sizes 450 and 750
Inquire
B32371
Holder for water tank for mounting on the rear of the appliance or the wall (connection lead 0.5m), additional water tank included, for HPP110 or HPP260 (standard equipment for HPP750)
Inquire
B04712
Central water supply
Inquire
GA1Q5
One year guarantee extension, size 160 and smaller