The Multifors 2 Cell is especially designed for multiple, parallel bioprocesses in small volumes. Together with the parallel bioprocess software INFORS Iris the Multifors 2 Cell offers a rapid and statistically accurate way to generate and understand process data.
The Multifors 2 Cell is offered in a choice of preconfigured packages and has an extensive list of options to optimize for a specific microbial fermentation and cultivation strategy.
Primarily used for Mammalian cells, Insect cells, Viruses or Vaccines and Cultures from 150 to 750 mL. Applications include process development, process optimization, scale up, scale down, statistical studies, process analytical technology (PAT), media development, growth studies toxicity testing and high throughput screening.
Precise control or monitoring of temperature, stirrer speed, pH, pO2, antifoam/level, feed, gas mix, gas flow, with 16 other channels free, for each vessel.
The design and construction of the vessels, with no welded parts, make the Multifors 2 Cell ideal for GMP applications.
Four high precision pumps per vessel are included for feed, acid, base and antifoam/level/harvest, ready for applications such as fed batch or continuous culture.
Features of the Multifors 2 Cell:
Faster results with optimized parallel handling
Flexible configurations designed for specific applications
Compact design, two bioreactors within a single base unit
Integrated OPC server for unlimited connection
Aseptic sampling
Clear path to scale up
Touch screen for precise control of up to 6 bioreactors
Eight fixed parameters per vessel, sixteen free
Two vessels with working volumes from 700 mL and 1000 mL with stainless steel top plate including many industry standard ports for a wide variety of sensors
Slow speed bottom drive magnetically coupled stirring system with speeds up to 300 rpm, pttched blade stirrer included as standard
Open frame gas supply with up to 5 mass flow controllers
Exit gas cooler, to minimize liquid loss
Four high precision pumps per vessel standard, simple removal and autoclaving with vessel without worrying about tubing
Aluminum heating block with integral cooling eliminates water jackets, allows good views of a bioprocess and precision temperature control
Specifications (Click here):
Bioreactors, the Benchtop Infors Multifors 2 Cell
Specifications
Model:
Multifors 2 Cell Bioreactor
Temperature range: (°C)
5 above coolant to 60
Stirrer speed: (rpm)
30 - 300
Capacity, base unit:
2 vessels
Maximum expansion:
3 base units per controller (6 vessels)
Working volume: (mL)
700, 1000, in pairs
Port fittings:
Stainless steel vessel top plate includes standard Pg13.5 ports for a wide variety of sensors such as antifoam, optical density, pH, pO2, redox or capacitance
Ports, 0.7 L Vessel: (Ø mm, n)
(12, 4), (10, 4), (7, 4) total 12
Ports, 1.0 L Vessel: (Ø mm, n)
(12, 5), (10, 4), (7, 4) total 13
Exit gas cooler:
Condenses water vapor to reduce loss
Pumps:
Four per vessel for feed, acid, base and antifoam/level/harvest
Gassing system:
Open frame, any combination of Air, O2, N2 or CO2 as a single gas or gas mix can be directed into the sparger or head space
Parameters:
Stirrer speed, temperature, pH, pO2, antifoam/level, feed, gas mix, gas flow and 16 free channels
Dimensions: W x D x H (mm)
350 × 500 × 850
Number
Description
Price $
MULTI-2-CELL
Multifors 2 Cell Individually Assembled Packages
Inquire
HT-GAS
INFORS HT Gas Analyzer, for exit CO2 and O2 gas analysis, optional multiplexer function for use with up to 6 bioreactors, immediate readings for culture analysis, precision control with Iris software, Iris software allows the calculation of parameters such as CO2 evolution rate, O2 uptake rate and respiratory quotient, as well as logging and storing data, analog connection allows use with non Infors bioreactors, compact 115 × 235 mm footprint, 142 × 275 × 135 mm overall size
Inquire
HT- SAFE
INFORS HT Super Safe Sampler, integrated non return valves, take smallest samples with no waste of culture, back flush with sterile air, sterile and safe sampling without laminar flow hood, sampling without risking contamination of the culture vessel, flushing of sample line not required (no dead volume), needle free, autoclavable many times
Inquire
BIO-IRIS
INFORS HT Iris software, flexible options for data logging, graphing and control, a link between all INFORS HT bioreactors, linked to Multifors, but PRO version is available for all types of bioreactors and includes more features such as links to balances or mass spectrometers, or communication features
Inquire
OPTIONS
Special optional solutions available for continuous, chemostat and perfusion processes, configured for batch and fed-batch processes by default
Inquire
BIO-SPIN
Spin filter for cell-free removal or harvesting of culture supernatants from bioreactors in continuous perfusion processes, cells are cultivated permanently under optimal conditions, filter is integrated into the device to give protection against contamination, available in various sizes and pore diameters