logo
Telephone: 513.942.4007 Rose Scientific Inc.
Toll Free: 888.249.4074
Fax: 780.462.5776
Email: usasales@rosesci.com
 
Search  
Quote Request Quote Request
Rapid Response Rapid Response
 
Home Products Service Clearance Chemicals Products COVID-19    
Rapid Response or call Rose Scientific toll free at 1.888.249.4074
Send us an email message outlining any request and we will get back to you with the details!
Contact me Send information Quote request  
Your name Your email Your telephone number

Climatic Test Chamber Model CTC 256 and Temperature Test Chamber Model TTC 256

  Overview

Rapid, precise and energy-saving temperature changes in the range from -42 ºC to +190 ºC, along with fast humidity recovery times, make the Memmert TTC temperature test chamber and the CTC climatic test chamber with humidity unit a perfect duo for controlled material testing, function tests, ageing tests or climate tests on composites, concrete, plastics or electronic components in material technology, the automotive and aerospace industries, and electronics.

To avoid condensation in the interior, the appliance has various heating systems, including low-voltage door frame and seal heating, contacted via door contacts, condensation water outlet pipe heating, and heated glass sheet and glass frame heating if optional glass door is selected.

The temperature is lowered through an integrated cooling unit. The automatic defrosting system for the cooling unit ensures the perfect running of the test chamber at low temperatures and in permanent operation.

The CTC climatic test chamber can in addition lower humidity in the chamber to 10 % rh or increase it to up to 98 % rh. Humidification takes place via a PTC hot steam generator. The amount of steam is controlled by two self-priming tube dosing pumps. Dehumidification takes place through controlled cooling the evaporator temperature below the dew point

Steel grids and silicone plugs to close up the cable feed-through in the right-hand side wall are provided.


Climate Test Chamber CTC 256 with optional laptop PC Control
Features:
  • Temperature range -42 °C to +190 °C
  • Compact size (256 L)
  • Electronically controlled active humidification and dehumidification from 10% to 98% rh
  • High-tech insulation system for permanent insulation capacity and energy efficiency
  • Average temperature change speed in heating operation: 10 K/min
  • Average temperature change speed in cooling operation from +190°C to 0°C: 6 K/min
  • Average temperature change speed in cooling operation according to IEC 60068-3-5: 4 K/min
  • Chlorine-free, ozone-neutral refrigerant R 404 A



CTC 256 Display
  • State-of-the-art control technology based on the Memmert Perfect performance class
  • Two 10 L tanks to collect condensation water are accessible in the drawer at base of unit
  • The CTC 256 is provided with two additional 10 L tanks to provide fresh water to humidify the interior


Specifications (Click here):  Climatic Test Chamber Model CTC 256 and Temperature Test Chamber Model TTC 256

As much function as necessary, as much individuality as possible! Contact Rose Scientific to custom design the Memmert Climatic Test Chamber or Temperature Test Chamber that does what it needs to do with the resources available.

The ordering number is assembled from the catalog numbers of the various options at the time the Test Chamber model is ordered. Generation 2003 options and accessories are available for the CTC 236 and TTC 236.

NumberDescriptionPrice $
CTC 256 Memmert Climatic Test Chamber Model CTC 256
Inquire
TTC 256 Memmert Temperature Test Chamber Model TTC 256
Inquire
B0 Full-sight glass door (5-layer insulating glazing), heated
Inquire
E20591 Stainless Steel Grid Inquire
V3 Temperature profile write/read unit for programming via PC with USB for writing to and reading from the chip card, up to 40 ramps
Inquire
V4 Additional chip card, blank, formatted (32 kb MEMoryCard XL for a maximum of 40 ramps) Inquire
V7 Chip card, pre-programmed for common test standards Inquire
V1 Oven-linked authorization card (User-ID-Card) - prevents undesired manipulation by unauthorized third parties Inquire
V2 Computer interface RS485 (for networking a maximum of 16 ovens) instead of interface USB Inquire
W6 RS232 interface instead of USB Inquire
V6 Connection cable for computer interface RS232 according to DIN 12 900-1 Inquire
W4 Interface Ethernet instead of RS232 inclusive software "Celsius Ethernet-Edition" Inquire
W1 Parallel/USB converter cable with integrated power supply unit to connect PCL3-compatible HP printers with USB interface to MEMMERT units. Inquire
W2 Documentation package consisting of parallel USB converter cable including PLC3-compatible HP color inkjet printer with USB interface (HP Deskjet 6980 or successor) for direct connection of printer to Memmert unit Inquire
W7 USB connection cable for computer interface Inquire
B04410 External control and logging package consisting of mini-Notebook and Software "Celsius", pre-configurated, and lateral swinging arm Inquire
H4 Flexible Pt100 for positioning in chamber or in load with the socket in the front control panel, 4-pin according to NAMUR NE 28 for external temperature recording (load temperature)
Inquire
H5 Potential-free contact (24V/2A) with socket 3 pin, according to NAMUR NE 28 for external monitoring (indication set point reached)
Inquire
H6 Potential-free contact (24V/2A) for combined fault message (e.g. supply failure, sensor fault, fuse) Inquire
H7 Floating triple contact, for signal generation, controlled by program segment (using PC), for a total of 3 freely selected functions to be activated (e.g. activation of audible and visual signals, exhaust motors, fans, stirrers etc.)
Inquire
D00110 Z4 Works calibration certificates at one temperature and humidity Inquire
D00103 IQ check list with works test data for oven as support for validation by customer Inquire
D00104 OQ check list with works test data for one freely selectable humidity and temperature value including temperature distribution survey for 27 measuring points to DIN 12880: 2007-05 as support for validation by customer Inquire
E05019 Software conforming to US FDA "Celsius FDA-Edition" for up to 16 units Inquire
FDAQ4 Integration of additional units (up to max.16 units) into an already existent US FDA software license Inquire